Main Features:
² Bandwidth : 50MHz/100MHz
² Measurement channel : 2 analog channels
² Real-time sampling rate : 500MS/s
² Storage depth : 64kpts
² Waveform capture rate : 5,000wfms/s
² Auto measurement : 34 waveform types
² Abundant trigger : Edge 、 Pulse width 、 Slope trigger 、 Video trigger 、 Alternating trigger
² Display : 7-inch WVGA ( 800 × 480 ) TFT LCD , super-widescreen, clean display
² Peripheral interface : Standard USB Host, USB Device, EXT Trig, Pass/Fail
² New A utoset function : Customized auto strategy to facilitate oscilloscope teaching
Quick Model Selection
Model |
UTD2052CL+ |
UTD2102CL+ |
Analog Bandwidth |
50MHz |
100MHz |
Channels |
2 |
2 |
Real-time |
500MS/s |
500MS/s |
Equivalence |
25GS/s |
25GS/s |
Storage depth |
64 kpts |
64 kpts |
Capture rate |
5000 wfms/s |
5000 wfms/s |
Rise Time (Typical) |
≤ 7ns |
≤ 3.5ns |
Technical Specification
Horizontal System Specification | |
Time-base scale |
2ns/div-50s/div |
Waveform interpolation |
Sin(x)/x |
Time-base accuracy |
≤ (50+2 × Service life)ppm |
Record length |
2 × 512k sampling point |
Storage depth |
Single channel: 64k ; Double channel: 32k |
Sampling rate and delay time accuracy |
± 50ppm (any time interval ≥ 1ms ) |
Measurement accuracy of time interval ( △ T) (full bandwidth) |
Single time: ± (1 sampling time interval+50ppm × reading+0.6ns) >16 average values: ± (sampling time interval+50ppm × reading+0.4ns) |
Vertical | |
Analog-to-digital converter (A/D) |
8bit |
Deflection factor range (V/div) |
1mV/div~20 V/div (at 1-2-5 increment) |
Position range |
≥± 8div |
Selectable bandwidth limitation (Typical) |
20MHz |
Low frequency response (AC Coupling , -3dB) |
≤ 5 Hz(above BNC) |
DC gain accuracy (sampling or average sampling mode) |
5mV ~20V/div :≤± 3% 1mV ~2mV/div :≤± 4% |
DC measurement accuracy (average sampling mode) |
When vertical position is 0 and N ≥ 16: ± (4% × reading+0.1div+1mV) and selects 1mV ~2mV/div; ± (3% × reading+0.1div+1mV) and selects 10mV ~ 20V/div; When vertical position is not 0 and N ≥ 16: ± (3% × (reading + vertical position reading) + (1% × vertical position reading)]+0.2div) The setting from 5mV/div to 200mV/div plus 2mV; the setting value from 200mV/div to 20V/div plus 50mV |
Measurement accuracy of voltage difference ( △ V) (average sampling mode) |
Under the same setting and environment conditions and after averaging the captured waveforms with a quantity of ≥ 16, the voltage difference ( Δ V) between any two points on the waveform: ± (3% × reading+0.05div) |
Trigger System Specifications | |
Trigger sensitivity |
≤ 1div |
Range of trigger level |
Interior : From the screen center ± 10div EXT :± 3V |
Trigger level accuracy (Typical) applicable for the signal with rising and falling time ≥ 20ns |
Interior :± (0.3div × V/div) (within ± 4 div from the screen center) EXT :± (6% setting value+40mV) |
Pre-trigger capacity |
Normal mode/scan mode, pre-trigger/delay trigger, the pre-trigger depth is adjustable. |
Hold-off range |
80ns~1.5s |
Set the level to 50% (Typical) |
Operate under the condition of input signal frequency of ≥ 50Hz |
Trigger mode |
AUTO, normal, single |
High-frequency holdoff |
Hold off signals over 80kHz |
Low-frequency holdoff |
Hold off signals below 80kHz |
Trigger mode | |
Edge |
Rise, fall, arbitrary edge |
Pulse width |
Pulse width term :>、<、 = |
Polarity : positive pulse width, negative pulse width | |
Pulse width range : 20ns~10s | |
Slope trigger |
Slope condition : Positive slope (>, <, within the scope) ; Negative slope (>, <, within the scope) |
Time : 20ns~10s | |
Video trigger |
Trigger sensitivity (Typical) : 2div Vpp |
Signal model and line/field frequency (video trigger type) : Support standard NTSC and PAL, and the line number scope is respectively 1-525 (NTSC) and 1-625 (PAL) | |
Alternating trigger |
Alter : Edge, Pulse, Slope |
Measurements | |
Cursor |
Manual mode Voltage difference between cursors ( △ V), Time difference between cursors ( △ T), Reciprocal of △ T (Hz) (1/ △ T)) |
Track mode : Voltage value and time value of point of waveform. | |
Auto measurement mode : Cursor display is allowed on auto measurement mode. | |
Automatic measurement |
Vpp, Vamp, Vmax, Vmin, Vtop, Vbase, Vmid, Average, Vrms, Overshoot, Preshoot, Frequency, Period, RiseTime, FallTime, +Width, Width, +Duty, Duty, Delay, FRFR, FRFF, FFFR, FFFF, FRLF, FRLR, FFLR, FFLF |
Measurement quantity |
Display 5 types of measurement at the same time. |
Measurement scope |
Screen or cursor |
Measurement statistics |
Average value, maximum value, minimum value and standard deviation. |
Math | |
Math operation |
+,-, × , ÷ |
Window |
Rectangle 、 Hanning 、 Blackman 、 Hamming |
Vertical scale |
Vrms 、 dBVrms |
Digital filtering |
Low pass, high pass, band pass, band reject |
Storage | |
Setting |
Internal: 20 groups. USB: 200 groups |
Reference waveform |
Internal: 20 groups. USB: 200 groups |
Data file |
Internal: 20 groups. USB: 200 groups |
Bitmap |
USB: 200 groups, in BMP format. |
Input Channel Specifications | |
Input Coupling |
DC, AC and GND |
Input impedance |
(1M Ω± 2%)//(18pF ± 3 pF) |
Probe attenuation coefficient |
0.01 × /0.02 × /0.05 × /0.1 × /0.2 × /0.5 × /1 × /2 × /5 × /10 × /20 × /50 × /100 × /200 × /500 × /1000 × |
Maximum input voltage |
400Vpk, the transient over voltage is 1000 Vpk. |
Display | |
Displays types |
LCD with Diagonal of 178mm (7-inch) |
Display resolution |
800 horizontal × RGB × 480 vertical pixels |
Display color |
Color |
Waveform luminance |
Adjustable |
Backlight intensity (Typical) |
300nit |
Language |
Multi-language |
Interface function | |
Standard configuration |
Standard USB Host, USB Devic e, EXT Trig, Pass/Fail Option : Multimeter module (UT-M12), LAN |
Trigger frequency meter | |
Reading resolution |
6bits |
Trigger sensitivity |
≤ 30Vrms |
Accuracy (Typical) |
± 51ppm (+1 character) |
Probe compensator output | |
Output voltage (Typical) |
About 3Vpp, when the load ≥ 1M Ω |
Frequency (Typical) |
10Hz,100Hz,1kHz (Default), 10kHz |
Power Source | |
Power voltage |
100V-240V~(Fluctuations 10% ) , 50/60Hz |
Power consumption |
100VA max |
Fuse |
F 1.6A 250V |
Environment Specifications | |
Intended use |
Indoor use |
Pollution degree |
2 |
Operating temperature |
Operating Temperature Range: 0 ℃ ~+40 ℃ |
Storage Temperature |
Storage Temperature Range: -20 ℃ ~+60 ℃ |
Cooling |
Build-in cooling fan |
Operating Humidity Range |
<35 ℃ : ≤ 90%RH 35 ℃ ~40 ℃ : ≤ 60%RH |
Operating Altitude |
Operating 2000 meters below Non-operating 15000 meters below |
Mechanical specifications | |
Size |
306mm(W) × 138(H) × 124 mm(D) |
Weight |
Excluding package:2.5kg Including package: 3kg |
Recommended calibration Interval | |
The recommended calibration interval is one year. |